Instructions for use Zeus Infinity Empowered SW 2.n 317
Fault – Cause – Remedy
Metering module faults
In operation
Fault Cause Remedy
No metering or too high/low
metering.
The metering module is empty. Refill the metering module.
No anesthetic agent concentra-
tion set at the device.
Set anesthetic agent concentra-
tion at the device.
No metering module connected Connect metering module.
The automatic anesthetic agent
detector indicates a different an-
esthetic agent from the metering
module.
Device error or incorrect filling.
Do not activate the metering
module!
Contact DrägerService
Smell of anesthetic agent, re-
lease of anesthetic agent va-
por.
Leakage at the interface or meter-
ing module faulty
(Self test failed.)
Do not activate the metering
module!
Change metering module.
Contact DrägerService about the
faulty metering module or if the
error message occurs again.
The lock of the filling device is not
tightened or the seal is defective.
Tighten the lock of the filling de-
vice firmly and check the seal.
Contact DrägerService if neces-
sary.
Drain plug is not closed.
(Not with DES)
Tighten drain plug.
The lever of the safety filling de-
vice is too slack so the seal is not
under sufficient pressure.
Have the lever readjusted.
Contact DrägerService
The locking bolt of the filling de-
vice is not completely inserted.
Loosen the lever, depress the
locking bolt completely, retighten
the lever.
The sight glass display is faulty. Contact DrägerService
Anesthetic gas scavenging sys-
tem AGSS is not connected
Connect the AGSS.
Anesthetic agent leaking. The locking bolt/cap is not locked. Lock it.