Instructions for use Zeus Infinity Empowered SW 2.n 15
For your safety and that of your patients
Product-specific safety information
WARNING
Risk of misinterpretation
Misdiagnosis or misinterpretation of the
measured values or other parameters may en-
danger the patient.
Do not make therapeutic decisions based
solely on individual measured values and
monitoring parameters. Therapeutic deci-
sions must be made solely by the user.
WARNING
Risk of malfunction
Unallowed modifications to the medical de-
vice lead to malfunctions.
This medical device may not be modified with-
out permission from Dräger.
WARNING
Risk of malfunction
Device failure or user errors may compromise
the correct functioning of the device. The
medical device does not react automatically to
certain changes in the patient condition, oper-
ating errors, or failure of components.
Continuously monitor the medical device so
that corrective measures can be initiated im-
mediately.
WARNING
Risk of device failures
The device may fail if the power supply is in-
terrupted.
Always connect the device on an uninterrupt-
ible power supply.
WARNING
Risk of not hearing the alarm tone
Dräger recommends the user to remain in the
vicinity of the anesthesia machine, i.e. within
a distance of up to 4 meters (12 ft). This facili-
tates fast recognition and response in the
event of an alarm.
WARNING
Risk of insufficient ventilation
Device failure or operating errors can lead to
ventilation failure.
To ensure immediate remedial action in
case of device failure, the device may only
be operated under permanent supervision
of users.
The general safety standards for anesthe-
sia systems require that a manual resusci-
tator be kept at the ready for emergency
ventilation.
WARNING
Risk of burns
Do not use humidifiers or heated hoses!