Instructions for use Zeus Infinity Empowered SW 2.n 105
Startup
Startup in emergencies
1 Touch the button (A) on the screen.
2 Set the rotary safety knob O
2 (B) for O2 emer-
gency delivery to the required O2 flow:
Range 0 to 12 L/min
3 Start manual ventilation.
Wait for the software to launch and for electronics
test to end. The device only performs a minimum
test.
After switching on, the device opens the Test de-
tails dialog window.
WARNING
Use only in urgent cases if there is no time for
the self test!
If a self test is not completed, malfunctions
can go undiscovered!
Pay special attention during operation!
050071
NOTE
During this phase, no anesthetic gas delivery is
possible!
O2
Air N2OO2 Air N2O
A
B
316
Test details
Self test
C