Instructions for use Zeus Infinity Empowered SW 2.n 409
Technical data
Measuring system
Pressure measurement (piezo-resistive)
Airway pressure, real-time curve Paw: –20 to 100 hPa
Measured values PEEP, PIP, Pplat; P
mean Measuring range: –20 to 100 hPa
Resolution: 1 hPa
Accuracy: ±4 % or ±2 hPa
Central supply pressure (CS) Measuring range: 0 to 9.8 kPa x 100
Resolution: 0.1 kPa x 100
Accuracy: ±0.1 kPa x 100
Cylinder pressure ("SilverLine" pressure reducer) Measuring range: 0 to 250 kPa x 100
Resolution: 1 kPa x 100
Accuracy: ±10 kPa x 100
Pressure gauge (mechanical) Measuring range: –20 to 80 hPa
Accuracy: ±5 mbar of the reading
Flow measurement (hot-wire anemometry)
Flow real-time curve Measuring range: –180 to 180 L/min
Resolution: 0.1 L/min
Accuracy at 60 L/min: ±10 % (under calibration con-
ditions)
Volume real-time curve Measuring range: –200 to 2000 mL
Resolution: 1 mL
Accuracy at 60 L/min: ±10 % (under calibration con-
ditions)
Tidal volume VT (inspiratory and expiratory) Measuring range: 0 to 4000 mL
Resolution: 1 mL
Accuracy:
<100 mL: ±10 % or ±15 mL (under calibration con-
ditions)
>100 mL: ±10 % (under calibration conditions)