EasyManua.ls Logo

Dräger Zeus Infinity Empowered - Adjusting Mixed Gas; Gas Delivery; Nitrous Oxide-Free System

Dräger Zeus Infinity Empowered
453 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Zeus Infinity Empowered SW 2.n 115
Operation
Gas delivery
Zeus IE can be operated in the manual fresh-gas
delivery mode (FG ctrl. control mode) and in the
closed control loop (Auto ctrl. control mode).
In contrast to conventional fresh-gas delivery, the
automatic control of oxygen, carrier gas and anes-
thetic agent takes place in a closed control loop.
Nitrous oxide-free system
A nitrous oxide-free device system can be config-
ured under Basic settings > More settings >
VA/N
2O (see page 350).
Throughout the system in nitrous oxide-free opera-
tion, N
2O monitoring and N2O setting elements are
turned off.
Adjusting mixed gas
1 Touch the MixGas standby button (A).
The MixGas settings dialog window containing the
control modes opens.
Switch control mode to:
FG ctrl. (Fresh-gas control) or
Auto ctrl. (Auto control)
WARNING
If face masks are used, the end expiratory an-
esthetic agent measurement may be incorrect
on account of the leak to be expected. Do not
operate the device in the Auto ctrl. control
mode.
WARNING
If even minute quantities of volatile anesthetic
agents have to be avoided for medical rea-
sons, e.g., malignant hyperthermia:
Remove DIVA metering modules from
Zeus IE.
Change filters and hoses.
Replace soda lime.
Change absorber filter in the breathing
system.
CAUTION
Mixed gas is only delivered if a ventilation mode is
active.
329
MixGas standby
A

Table of Contents

Related product manuals