Cleaning, disinfection, and sterilization
380
Instructions for use Zeus Infinity Empowered SW 2.n
3 Slide the new disposable CO2 absorber on the
holder (D).
4 Press the CO
2 absorber into the device (D) until
it engages. See also: page 67.
Attaching the breathing hoses
For operation in a closed system, hose water traps
can be used in order to collect the condensate (at-
tach at the lowest point).
1 Attach the breathing hoses (A) to the inspiration
and expiration nozzles of the breathing system
and connect to the Y-piece.
2 Attach the HME filter (B) or filter to the nozzle of
the Y-piece.
3 Establish the Luer-Lock connection (C) of the
sample line between the HME filter of the Y-
piece and the CO2 water trap.
4 Connect the breathing bag (D) with the connec-
tion nozzle to the breathing hose. Attach the
shorter breathing hose to the angled nozzle and
attach the breathing bag to the bag holder so
that it hangs down.
For perfect operation in a closed system, the use of
Dräger breathing bags (see list of accessories) is
recommended.
Assembling the anesthetic gas receiving
system (AGS)
1 Install particle filter, tighten the union nut (A).
2 Screw on the flow tube (B).
3 Tighten the union nut (C).
4 Reattach the buffer-volume container (D).
WARNING
Check the breathing hoses for damage before
each use. Do not use damaged breathing
hoses.
028
A
A
B
C
C
D
WARNING
Do not use any antistatic or conductive
breathing hoses!
Risk of burns when using HF surgery equip-
ment.
NOTE
Zeus IE is manufactured without natural rubber
latex.
To minimize the risk of exposure to latex, use
breathing bags and breathing hoses manufac-
tured without natural rubber latex.
029
•
A
B
C
D