Assembly and preparation
74
Instructions for use Zeus Infinity Empowered SW 2.n
DIVA metering modules for anesthetic agents
The device can be operated with one or two DIVA
metering modules. The selection of the slot is arbi-
trary.
z Fill the DIVA metering module with the appropri-
ate anesthetic agent.
Filling can be done with the DIVA metering
module plugged in or outside of the device.
Filling is also possible during anesthesia.
Before the internal self test of the device, the
DIVA metering modules must be filled and
connected in order to ensure a complete
test.
Inserting and removing the DIVA meter-
ing module
The DIVA metering module can be exchanged in all
operating states (switched off, on standby or during
anesthesia).
Plugging in the DIVA metering module
z Insert the DIVA metering module into the slot
until it can be felt and heard to have engaged.
Removing the DIVA metering module
z Keep the unlock button (A) pressed and remove
the DIVA metering module.
WARNING
Storing the DIVA metering module
only when empty (except if supplied for im-
minent clinical use)
in well-ventilated rooms
–not over 40°C
Empty the DIVA metering module for periodic
maintenance and for transportation!
040
NOTE
Pressing the unlock button during operation re-
sults in immediate shutdown of the DIVA metering
module! This can lead to a rapid drop in anes-
thetic agent concentration in the breathing gas.
A
A