Cleaning, disinfection, and sterilization
366
Instructions for use Zeus Infinity Empowered SW 2.n
Removing the endotracheal suction sys-
tem (optional)
1 Detach the suction hose (A) to the endotracheal
suction hose.
2 Detach the vacuum hose (B).
3 Grasp the bottle cap by the silicone sleeve (D)
and pull it off.
4 Take the secretion collection bottle and the rins-
ing bottle (E) out of the holder and empty them.
5 Remove the filter (C) from the base of the en-
dotracheal suction unit. Remove the vacuum
hose.
Change the filter every 14 days.
6 Unbutton the silicone sleeve (F) from the cover
of the secretion collection bottle.
7 Pull the overflow protection float (G) out of the
riser pipe.
When using the disposable VacuSmart container:
Dispose of the VacuSmart container and hose con-
taining the infectious secretion according to regula-
tions.
033024
A
B
C
D
E
E
F
G
WARNING
Risk of infection
The contents of the suction bottle can be
highly infectious.
– Wear protective gloves when emptying the
suction container.
– Follow the hospital hygiene regulations.