Operation
140
Instructions for use Zeus Infinity Empowered SW 2.n
IV system – IVenus
Information and safety instructions on
the IV application
Operation of the Module DPS syringe pump and
other safety instructions are explained in separate
instructions for use.
Attach IVDock and Module DPS to Zeus IE,
see page 73. Note the respective instructions for
use before using these modules.
The Module DPS syringe pump can be remotely
controlled with IVDock on Zeus IE. Module DPS
may also be operated independently.
Remote control via Zeus IE is not possible until op-
eration of the Module DPS syringe pump is termi-
nated. The control panels for the IV system on the
screen remain gray for that period of time (i.e. they
are not operable).
If there is an interruption in the data link between
Zeus IE and IVDock, the pumps cannot be re-
motely controlled. The syringe pumps then con-
tinue running at the set metering and can be con-
trolled independently of Zeus IE.
If remote control of a syringe pump is not possible,
the syringe pumps must be reconfigured.
Observe the Module DPS instructions for use:
1 Activate configuration group PAr.
2 Set syringe selection PAr 4.
3 Set at least one syringe type to SEL.
If any malfunctions should occur on Zeus IE (IVe-
nus), the syringe pumps can be controlled inde-
pendently of Zeus IE:
z Remove the IVDock plug from the IV socket on
Zeus IE. Observe the instructions for use of the
Module DPS syringe pump!
WARNING
Risk of patient injury
Synchronous or parallel and overlapping op-
erations of the pump and Zeus IE may lead to
operating errors.
The current operations must be completed or
interrupted, before new operations are under-
taken with other devices (Module DPS or
Zeus IE).
WARNING
Risk of patient injury
When starting the Module DPS syringe pump,
the syringe size, syringe type and drug name
must be confirmed directly on the Module DPS
syringe pump. To eliminate the risk of incor-
rect settings and ensure the patient's safety,
the user must check must check on the sy-
ringe pump and on Zeus IE that the correct
drug is being used with the correct dose.
WARNING
Risk of patient injury
Only use weight-related metering from a pa-
tient weight of 5 kg upwards.
Correct adjustment of patient weight is re-
quired because otherwise in the event of
weight-based metering the drug may be me-
tered incorrectly.
NOTE
When the IVDock plug is disconnected from the
IV socket on Zeus IE, the base (IVDock) is no
longer supplied with power. Each Module DPS is
powered by its own battery. Pay attention to the
battery warning! Infusion is automatically stopped
when the battery is empty.