Cleaning, disinfection, and sterilization
370
Instructions for use Zeus Infinity Empowered SW 2.n
Sterilization
Sterilization eliminates living microorganisms from
semi-critical medical devices and dries residual wa-
ter in the interior of components.
z Sterilize only components that have been
cleaned and disinfected.
For sterilization, use a vacuum steam sterilizer (in
accordance with DIN EN 285), preferably with frac-
tional vacuum.
Sterile accessories
WARNING
Risk of fire
Residual vapors of highly flammable disin-
fectants (e.g., alcohols) and deposits that
were not removed during reprocessing may
ignite when the flow sensor is in use.
Ensure particle-free cleaning and disinfec-
tion.
After disinfection, allow the flow sensor to
air for at least 30 minutes.
Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
cles.
Replace flow sensors when damaged,
soiled, or not particlefree.
CAUTION
Risk of failure of flow measurement
Improper reprocessing and soiling, such as de-
posits or particles, may damage the flow sensor:
No machine cleaning or disinfection of the
sensor insert
No plasma sterilization or radiation steriliza-
tion
No compressed air
No water jets, compressed air, brushes, or
similar to be used on the sensor insert
No ultrasonic bath
Clean and disinfect the flow sensor in accord-
ance with the corresponding instructions for
use.
For disinfecting the flow sensor use only clean
disinfectant solutions.
CAUTION
These medical devices have been designed,
tested, and manufactured exclusively for single
use. Disposable products must not be reused, re-
processed, or sterilized. Reuse, reprocessing, or
sterilization may lead to failure of the medical de-
vices and injury to the patient.
CAUTION
Do not used sterile packed accessories if the
packing has been opened or damaged or there
are other indications of non-sterility. Reprocess-
ing and re-sterilization of disposable articles are
not permitted.
WARNING
Do not allow any alcohol or alcoholic agents
to get into the sample line openings! Alcohol
may distort the measuring results.