Alarms
174
Instructions for use Zeus Infinity Empowered SW 2.n
Adjusting the lower and upper limit val-
ues according to set points and patient
demographic data
Alarm limit Algorithm Remark Default values (Range) for
Adult
2
Adult
1
Ped2 Ped1 Neo1
Autoset IBP:
ART S
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
ART D
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
ART M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
LV S
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
LV D
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
LV M (mmHg) see page 173 40 %
(OFF, 20 to 80)
GP1 S, GP2 S
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
GP1 D, GP2 D
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
GP1 M, GP2 M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
PA S (mmHg) see page 173 40 %
(OFF, 20 to 80)
PA D (mmHg) see page 173 40 %
(OFF, 20 to 80)
PA M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
RV S
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
RV D
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
RV M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
CVP M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)