Instructions for use Zeus Infinity Empowered SW 2.n 175
Alarms
LA M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
RA M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
ICP M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
Autoset Hemodynamics:
ART S
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
ART D
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
ART M
(mmHg)
see page 173 40 %
(OFF, 20 to 80)
HR/PLS see page 173 40 %
(OFF, 20 to 80)
SpO
2 see page 173 A fixed value is set 100 %
(fixed)
98 %
(fixed)
SpO
2 current value – X %
absolute
absolute Autoset 5 % absolute
(OFF, 2 to 10)
NIBP S see page 173 40 %
(OFF, 20 to 80)
NIBP D see page 173 40 %
(OFF, 20 to 80)
NIBP M see page 173 40 %
(OFF, 20 to 80)
T1 see page 173 7 %
(OFF, 5 to 20)
T2 see page 173 7 %
(OFF, 5 to 20)
TBlood see page 173 7 %
(OFF, 5 to 20)
Autoset ventilation/anesthesia:
inN
2O see page 173 40 %
(OFF, 20 to 80)
inHal see page 173 40 %
(OFF, 20 to 80)
Alarm limit Algorithm Remark Default values (Range) for
Adult
2
Adult
1
Ped2 Ped1 Neo1