Technical data
408
Instructions for use Zeus Infinity Empowered SW 2.n
Fresh-gas outlet for non-rebreathing system (optional)
Connection ISO cone (male), 22 mm diameter with ISO cone
(female), 15 mm diameter
Pressure Max. 80 hPa at 18 L/min
Fresh-gas flow 3 to 18 L/min
For tolerances see fresh-gas delivery
Airway resistances
Airway resistances, in a dry state without patient hoses (insp. + exp.)
Inspiratory Expiratory
In accordance with ISO 8835-2 at a flow of
60 L/min (depending on the configuration of the
breathing system, both the insp. and exp. value
may deviate by ±0.3 hPa)
1.2 hPa 5.5 hPa
289
Pinsp (hPa) with standard
absorber
P
exp (hPa) with standard
absorber
Flow (L/min)
Pressure(hPa)