Operating concept
48
Instructions for use Zeus Infinity Empowered SW 2.n
Color concept
Therapy controls
Set values or operating states
1 Touch the therapy control. The color turns yel-
low.
2 Turn the rotary knob to set the value.
3 Press the rotary knob to confirm the value. The
color of the therapy control turns dark green or
bright green.
In the following chapters of the instructions for use,
this sequence of action is simply written: "Use the
rotary knob to set and confirm the value."
Waveforms and parameters
The screen setup is used to set the colors for the
waveforms and measured values, see page 334.
Color coding for anesthetic agents and
medical gases
Standardized color coding in accordance with
ISO 5359 / ISO 32 / ISO 5360 is used to identify an-
esthetic agents and medical gases.
The colors for O2, Air, and N2O are adapted to lo-
cally applicable standards.
306
Gray Non-operable element.
Dark gray Currently not operable, function ac-
tivated
Yellow Selected function, not yet confirmed
with rotary knob.
Light green Operable element, function not acti-
vated.
Dark green Operable element, function acti-
vated.