EasyManua.ls Logo

Dräger Zeus Infinity Empowered - Direct Suspension of Alarms Using Buttons; In the Main Menu Bar; Suspending Alarms

Dräger Zeus Infinity Empowered
453 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
166
Instructions for use Zeus Infinity Empowered SW 2.n
Suspending alarms
Alarm groups can be switched off and on.
z Touch the Suspend tab (A).
The following alarm groups can be suspended:
If the cardiac bypass mode is activated, the follow-
ing alarm groups are switched off:
Apnea alarms
–CO2 alarms
NIBP/BP alarms
–SpO2 alarms
Heart rate alarms
ST alarms
Arrhythmia alarms
Technical alarms
Technical alarms can be acknowledged and then
set to a low priority.
Acknowledging a technical alarm:
z Touch the Confirm button (G).
The respective alarms are displayed as alarms of
lower priority.
Direct suspension of alarms using but-
tons in the main menu bar
The suspension of an alarm group can be config-
ured for direct access as a button in the main menu
bar. See "Configuring buttons" on page 332.
CO2 alarms On/Off
The CO
2 alarms off button can be used to switch
off the alarm limits for inCO2 and etCO2.
The symbol is displayed.
z Pressing the CO
2 alarms off button again
switches the CO
2 alarms on again.
Alarms are activated after detection of a respiratory
phase.
If still no respiratory phase has been detected after
changing from standby to a ventilation mode, the
CO
2 alarm is switched off temporarily. The sym-
bol is displayed. The symbol disappears as soon as
respiratory phases are detected, see CO
2 auto
wake-up (breaths) page 349.
363
Alarm group Factory setting
Apnea alarms (B) On
CO
2 alarms (C) On
BIS/NMT alarms (D) On
CV alarms (E) On
Cardiac bypass mode (F) Off
Alarms
A
B
C
D
E
F
G
CAUTION
Suspending technical alarms does not eliminate
the cause of the alarm. Eliminate the cause of the
alarm without delay, see page 255!

Table of Contents

Related product manuals