Assembly and preparation
76
Instructions for use Zeus Infinity Empowered SW 2.n
Transportation of filled DIVA metering
modules
The DIVA metering module can be transported in
any position. The filling device must be properly
closed.
Transportation in filled state
– Only within the scope of regular clinical opera-
tion, not for storage and dispatch
– Only with ambient conditions in accordance
with the technical data
– The DIVA metering module may remain filled
provided operating interruptions do not exceed
6 months; otherwise, empty the DIVA metering
module.
The anesthesia machine can be moved around the
workplace when the DIVA metering module is
plugged in and locked.
WARNING
Risk of patient injury
Do not use anesthetic agents that were emp-
tied from a DIVA metering module, but dispose
of in a proper manner. Otherwise, the anes-
thetic agents may be contaminated, mixed
with others or used by mistake.
CAUTION
Increased occupational exposure
When emptying the DIVA metering module, do
not fill the anesthetic agent bottle up to the brim!
This can cause anesthetic agent to be released
into the atmosphere.
CAUTION
Health risk
Fill the DIVA metering modules only up to a tem-
perature of 40 °C, otherwise anesthetic agent can
escape.
CAUTION
Increased occupational exposure
When filling during the warm-up phase of the
DIVA metering module, anesthetic agent may es-
cape.
Only fill the DIVA metering module after the warm-
up phase.