Technical data
404
Instructions for use Zeus Infinity Empowered SW 2.n
Gas delivery, control mode FG ctrl.
Concentration of volatile anesthetic agents in the
fresh gas
Isoflurane 0 to 5 Vol.%SL (max. 1000 mL/min saturated steam)
Accuracy: ±20 % or 0.2 Vol.%
SL
Resolution: 0.1 Vol.%SL
Sevoflurane 0 to 8 Vol.%SL (max. 1000 mL/min saturated steam)
Accuracy: ±20 % or 0.2 Vol.%
SL
Resolution: 0.1 Vol.%SL
Desflurane 0 to 18 Vol.%SL (max. 1500 mL/min saturated
steam)
Accuracy: ±20 % or 0.2 Vol.%SL
Resolution: 0.2 Vol.%SL
Gas delivery, control mode Auto control
Exp. concentration of volatile anesthetic agent (With automatic control)
Isoflurane 0 to 2.5 Vol.%
SL (max. 1000 mL/min saturated
steam)
Accuracy: ±20 % or 0.2 Vol.%SL
Resolution: 0.1 Vol.%SL
Sevoflurane 0 to 5 Vol.%SL (max. 1000 mL/min saturated steam)
Accuracy: ±20 % or 0.2 Vol.%
SL
Resolution: 0.1 Vol.%SL
Desflurane 0 to 12 Vol.%SL (max. 1500 mL/min saturated
steam)
Accuracy: ±20 % or 0.2 Vol.%SL
Resolution: 0.2 Vol.%SL