Assembly and preparation
62
Instructions for use Zeus Infinity Empowered SW 2.n
Electrical fuse for auxiliary sockets
If an automatic fuse trips:
1 Remedy the malfunction.
2 Then switch on switch (A) on the automatic
fuse.
The fuse is active again.
CAUTION
For devices with power input with 8 A electronic
main fuse:
In the first 2 minutes after activating the Zeus IE,
no power may be drawn on the auxiliary sockets
because the total current fuse of the Zeus IE may
otherwise trip and the device may switch to bat-
tery operation.
If the main fuse trips, the alarm Mains supply
failure appears. In this case, wait for the power-
on procedure and then reactivate the main fuse.
Put connected devices into operation only after
this time.
CAUTION
To ensure electrical safety, only connect devices
to the serial ports (COM 1, COM 2, or COM 3), or
devices and networks to the network interface
(Ethernet connection), which have a maximum
nominal voltage of 24 V
DC and which meet one of
the following standards:
– IEC 60950-1: Ungrounded SELV circuits
– IEC 60601-1 (as of 2nd edition): Exposed sec-
ondary circuits