Configuration
344
Instructions for use Zeus Infinity Empowered SW 2.n
Infusion
The page Infusion > General (A) opens.
For all syringe pumps, the standard alarm limit can
be configured for each syringe size:
B Pressure
z Touch the corresponding button and set the de-
sired value.
The standard alarm limits are automatically acti-
vated if the drug is changed during operation. For
setting ranges, see chapter "Alarms > All limits > In-
fusion" on page 171".
Drug library
The drug data is managed in the drug library.
z Touch the Drug library tab (C).
The Access code dialog window opens. The 4-
digit access code can be entered.
The full description, see "Drug library"
on page 149.
Event recorder
The Event recorder > User log page opens.
All setting changes and events are documented.
Use the buttons (A) and (B) to scroll up and down
in the log book.
Use the Refresh button (C) to display the last event
in the log book.