Alarms
168
Instructions for use Zeus Infinity Empowered SW 2.n
All limits
Settings of all alarm limits
Alarm limits not currently required can also be set.
z Touch the All limits tab (A).
Alarm limits can be set for the following areas:
Hemodyn. (optional)
IBP (optional)
More IBP (optional)
Arrhythmia (optional)
Anesth.
Infusion (optional)
1 Touch the tab (B) of the respective area.
2 Touch the button of the alarm limit to be modi-
fied. The button turns yellow.
3 Set the required value by turning the rotary
knob and push to confirm.
Hemodynamics
z Touch the Hemodyn. tab (A).
The alarm limits of the following parameters can be
configured:
B HR (heart rate)
C SpO
2 (blood oxygen saturation)
D NIBP S (systolic blood pressure)
E NIBP D (diastolic blood pressure)
F NIBP M (mean blood pressure)
G T1 (temperature, measuring point 1)
H T2 (temperature, measuring point 2)
I TBlood (blood temperature)
J Brady (bradycardia)
Setting the detection threshold for bradycardia
(only in the Neonates and Pediatrics patient cat-
egories with ext. arrhythmia off)
Factory setting: Off; 20 to 100 1/min
K ST LAT (ST lateral)
L ST INF (ST inferior)
M ST ANT (ST anterior)
N NMT (TOF count)
O BIS (BIS index)
365
A
B
Alarms
365
Alarms
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O