Assembly and preparation
80
Instructions for use Zeus Infinity Empowered SW 2.n
DIVA metering module with safety filling device
for isoflurane
The safety filling system consists of the following
components:
A anesthetic-specific filling device on the DIVA
metering module
B anesthetic-specific Dräger filling adapter
C anesthetic-specific collar and threads on the
bottle neck
Recommendations
Use only anesthetic agent bottles with an anes-
thetic-specific collar on the bottle neck.
Use only Dräger filling devices.
Use filling adapters with a non-return valve only.
Use undamaged filling adapters and anesthetic
agent bottles only.
If the filling adapter is not firmly and tightly con-
nected to the anesthetic agent bottle and the
DIVA metering module, the anesthetic agent
can flow out.
When filling during operation
1 Touch the Ventilate reservoir (A) button before
filling.
The DIVA metering module is ventilated for ap-
prox. 30 sec. No anesthetic agent vapor es-
capes into the environment during filling.
2 Select the filling adapter for the appropriate an-
esthetic agent.
The color coding and names/symbols on the fill-
ing adapter correspond to the anesthetic agent
used.
3 Screw the filling adapter (D) firmly onto the an-
esthetic agent bottle.
051
A
B
C
332052
MixGas settings
A
D
E