Instructions for use Zeus Infinity Empowered SW 2.n 135
Operation
Changing the patient from the ventilator
standby
1 Switch to Vent. Standby.
Prerequisite: The New patient button has been
configured, see page 332.
2 Touch the New patient button (C).
The startup dialog opens.
3 Perform configuration, see page 109.
Checking the soda lime
Check the soda lime in the following cases:
– Soda lime in the absorber is violet.
– Inspiratory CO
2 concentration inCO2 is
5 mmHg or more.
z Replace soda lime, see page 66.
Changing the HME filter or filter
The filter on the Y-piece must be changed after
each patient. Connect the sample line again if nec-
essary.
z Remove old filter.
z Attach new filter firmly.
312473
Startup dialog
C
NOTE
Observe hospital regulations for infectious pa-
tients! Dispose of the old filter according to hospi-
tal hygiene regulations.