Operation
136
Instructions for use Zeus Infinity Empowered SW 2.n
Leak test
Checking all breathing gas-conducting compo-
nents for leaks
Determining compliance for volume correction
Calibration of the flow sensors
The leak test can only be performed in system
standby.
Perform leak test, if:
the breathing system was opened with the twist
locks
the soda lime was replaced
the breathing filters were replaced
the sample line was replaced
the breathing hoses were replaced
the flow sensors were replaced
Starting the leak test
1 Touch the Device test button (A).
WARNING
Risk of patient injury
Do not perform leak test when a patient is con-
nected! In this operating state, it is not possi-
ble to monitor the patient. During the test,
pressures and concentrations may occur in
the system which are hazardous to the pa-
tient!
CAUTION
After each change of patient, a leak test must be
performed to rule out the possibility of major
leaks.
048
NOTE
Before starting the leak test, plug the filter onto the
self-test adapter and connect the sample line.