Operation
136
Instructions for use Zeus Infinity Empowered SW 2.n
Leak test
– Checking all breathing gas-conducting compo-
nents for leaks
– Determining compliance for volume correction
– Calibration of the flow sensors
The leak test can only be performed in system
standby.
Perform leak test, if:
– the breathing system was opened with the twist
locks
– the soda lime was replaced
– the breathing filters were replaced
– the sample line was replaced
– the breathing hoses were replaced
– the flow sensors were replaced
Starting the leak test
1 Touch the Device test button (A).
WARNING
Risk of patient injury
Do not perform leak test when a patient is con-
nected! In this operating state, it is not possi-
ble to monitor the patient. During the test,
pressures and concentrations may occur in
the system which are hazardous to the pa-
tient!
CAUTION
After each change of patient, a leak test must be
performed to rule out the possibility of major
leaks.
048
NOTE
Before starting the leak test, plug the filter onto the
self-test adapter and connect the sample line.
300
Standby
Standby
A