Instructions for use Zeus Infinity Empowered SW 2.n 91
Startup
Checking the O2 flush
The O2 flush is available for flushing and filling the
breathing system and the breathing bag with oxy-
gen quickly (approx. 35 L/min), bypassing anes-
thetic agent delivery.
There is a knob (A) on each side of the breathing
system with identical functions.
1 Manually close the filter on the Y-piece or con-
nect the test lung.
2 Press the O
2+ knob (A).
3 The breathing bag fills; the filling process can
be heard.
Checking the O2 emergency delivery
1 Manually close the filter on the Y-piece or con-
nect the test lung.
2 Press O
2 rotary safety knob (A) to unlock it and
adjust O2 flow by turning.
3 The breathing bag fills.
4 Turn the O
2 rotary safety knob (A) back to the
original position and push it in.
Checking the breathing system
1 It is complete and engaged. The hoses are
firmly connected.
2 New soda lime, not discolored violet.
038042
AA
A
047
CAUTION
Risk of device failure
To avoid a backflow of condensed water into the
breathing system, the hoses between the patient
and breathing system must have sufficient slack.
The condensate in the hose must be continuously
monitored and, if necessary, disposed of or dis-
charged to the hose water traps. Standing water
impairs device operation!