Instructions for use Zeus Infinity Empowered SW 2.n 79
Assembly and preparation
Emptying the DIVA metering module
1 Hold empty anesthetic agent bottle (A) under
the drain opening on the bottom of the DIVA me-
tering module.
2 Screw the locking cap (B) off the filling opening.
3 Use a screwdriver for hexagon sockets to open
the drain plug (C).
4 Close the drain valve again after emptying.
5 Close the anesthetic agent bottle with the cap.
6 Close the DIVA metering module with the lock-
ing cap (B).
263
WARNING
Risk of patient injury
Do not use anesthetic agents that were emp-
tied from a DIVA metering module, but dispose
of in a proper manner. Otherwise, the anes-
thetic agents may be contaminated, mixed
with others or used by mistake.
A
B
C