Instructions for use Zeus Infinity Empowered SW 2.n 405
Technical data
Maximum inspiratory concentration of volatile an-
esthetic agents
(inspiratory limit with automatic control)
Isoflurane 0 to 5 Vol.%
SL
Accuracy: ±20 % or 0.2 Vol.%SL
Resolution: 0.1 Vol.%SL
Sevoflurane 0 to 8 Vol.%SL
Accuracy: ±20 % or 0.2 Vol.%SL
Resolution: 0.1 Vol.%SL
Desflurane 0 to 18 Vol.%SL
Accuracy: ±20 % or 0.2 Vol.%SL
Resolution: 0.2 Vol.%SL
Ventilation
Electronically controlled radial-flow compressor in
conjunction with electrically controlled proportional
valve, fresh-gas decoupled.
Ventilation modes Volume Control
Volume Control AutoFlow, with/without Sync.(VC)
Pressure Control, with/without Sync.(PC)
CPAP
MAN/SPON
Optional:
Pressure Support
. Support for Volume Control
AutoFlow, Pressure Control and CPAP
Pressure limitation Pmax for Volume Control and
Volume Control AutoFlow
(PEEP + 5) or (PEEP + Psupp) to 70 hPa
Accuracy: ±10 % or ±2 hPa
Resolution: 1 hPa
Inspiration pressure P
insp
for Pressure Control
PEEP to 70 hPa
Accuracy: ±10 % or at least ±2 hPa
Resolution 1 hPa
Gas delivery, control mode Auto control (Cont.)