Instructions for use Zeus Infinity Empowered SW 2.n 373
Cleaning, disinfection, and sterilization
Semicritical medical devices
Items which can be re-
processed
Recommended
reprocessing in-
tervals
Prelimi-
nary
cleaning
Machine
cleaning and
disinfection
Manual Sterilization
Cleaning Disinfec-
tion
Breathing system: Weekly Yes Yes Possible Possible Possible
– Breathing system
housing (upper
and lower part)
– Inspiratory/expira-
tory ports, APL
valve
– Inspiratory valve,
expiratory valve
– Bag elbow
– Breathing hoses After each pa-
tient
Observe the associated instructions for use.
Absorber container
and absorber insert
Weekly Yes Yes Possible Possible Possible
Soda lime dust filter Replace after
each refilling of
soda lime.
No No No No No
Blower module
TurboVent 2
As needed No No Only
outside
Only out-
side
Yes
Minimum of
60 sterilization
cycles
Sample line Replacement
only
– When the sample
line is fitted to the
filter on the Y-
piece.
Daily
– When the sample
line is fitted di-
rectly to the Y-
piece and the fil-
ters are fitted to
the breathing sys-
tem.
After each pa-
tient
No No No No No
Breathing system
mount
Weekly Yes No Only
outside
Only out-
side
No
AGS Weekly Observe the associated instructions for use.