Instructions for use Zeus Infinity Empowered SW 2.n 377
Cleaning, disinfection, and sterilization
SpO2 sensors
Rub down reusable sensors with a cloth soaked in
a soapy water. To disinfect, the sensor may be
rubbed down with a cloth moistened in diluted alco-
hol (70 %). Dry thoroughly with a lint-free cloth. The
sensor must be completely dry before the next use.
Temperature sensors
Clean sensor with hydrogen peroxide (H2O2; 3 %)
or alcohol (70 %) and disinfect with an aldehyde-
based disinfectant.
Sterilize the sensor with ethylene oxide. In so do-
ing, the thermistor must not come into contact with
liquid ethylene oxide.
During gas sterilization, a temperature of 45 °C
must not be exceeded. Degas the sensors after
sterilization and allow them to air-dry for 24 hours.
Trident pod, BISx pod
Do not allow any liquids to get into the pods. Only
rub down with a cloth soaked in soapy water.
Do not autoclave the pods!
CAUTION
These cleaning instructions apply to reusable
sensors only.
CAUTION
Do not irradiate or autoclave sensors or cables or
immerse them in water or other liquids. It can
cause malfunctions.
CAUTION
Do not use any disinfectants containing phenol;
disinfectants can be absorbed by vinyl plastics.
Do not use any highly aromatic agents or any
chlorine, acetone, ether or ester compounds!