Technical data
Measuring ranges T NMT: 20 to 40 °C
Single twitch: 0 to 200 % of the reference pulse
TOF Ratio: 0 to 150 %
TOF Count: 0 to 4 (number of stimulations) if TOF
Ratio <20 % or if the number of stimulations <4
PTC: 0 to 20 (number of stimulations)
Accuracy T NMT: ±1 °C
Single twitch: ±10 %
TOF-Ratio: ±10 %
Display resolution T NMT: 1 °C
Single twitch: 1 %
TOF-Ratio: 1 %
TOF-Count/PTC: 1
Trend display ranges Single twitch: 0 to 200 %, 5 % increments
TOF Ratio: 0 to 150 %, 5 % increments
TOF Count: 0 to 4, 1-unit increments
PTC: 0 to 20, 1-unit increments
Electrical specifications Type CF as per IEC 60601-1
Defibrillation protection according to IEC 60601-1A2
Associated standards IEC 60601-2-40; specification for electromyo-
graphs and devices for evoked potentials