EasyManua.ls Logo

Dräger Zeus Infinity Empowered - Gas Flow Diagram

Dräger Zeus Infinity Empowered
453 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Description
440
Instructions for use Zeus Infinity Empowered SW 2.n
Gas flow diagram
A Non-return valve to anesthetic gas scavenging
system (AGSS)
B APL valve
C PEEP/P
MAX valve
D Expiratory valve
E Pressure sensor airway pressure
F SpiroLife expiration flow sensor
G Patient's lung
H SpiroLife inspiration flow sensor
I Inspiratory valve
J CO
2 absorber
K Option CO
2 Clic absorber, unlocked position
L Safety valve in breathing system (excess pres-
sure)
M Inspiration pressure sensor
N Pressure gauge
O Additional air valve
P Blower
Q Pressure sensor – breathing bag
R Breathing bag
S Microbial filter sample gas return line
T Gas analysis
U CO
2 water trap
V External area for the leak test: for manual venti-
lation
W Closed circulatory system valve
X Man/Spont changeover valve
250
P
P
P
M
J
K
L
M
N
O
P
Q
R
S
T
U
V
W
X
Man
Spont
Fresh gas/AGas
O
2 flush/emergency-O2
A
B
C
D
E
F
G
H
I

Table of Contents