Instructions for use Zeus Infinity Empowered SW 2.n 431
Technical data
Electromagnetic environment
The medical device is intended for use in an elec-
tromagnetic environment as specified in the follow-
ing. The user must ensure that the medical device
is used in such an environment.
Transmissions Compliance
with
Electromagnetic environment
Radio frequency emissions
(CISPR 11)
Group 1 The medical device uses radio frequency
energy only for its internal function. There-
fore, its radio frequency emissions are very
low and are not likely to cause any interfer-
ence in nearby electronic equipment.
Class A The medical device is not specified for op-
eration in residential buildings and such
equipment as is directly (without trans-
former) connected to the same low voltage
network as the residential building.
Harmonic emissions
(IEC 61000-3-2)
Not applicable
Transmission of voltage
fluctuations / flicker
(IEC 61000-3-3)
Not applicable