Instructions for use Zeus Infinity Empowered SW 2.n 375
Cleaning, disinfection, and sterilization
SpiroLife flow sensor
Reprocess the flow sensor in accordance with the
corresponding instructions for use.
CAUTION
Risk of failure of flow measurement
Improper reprocessing and soiling, such as de-
posits or particles, may damage the flow sensor:
No machine cleaning or disinfection of the
sensor insert
No plasma sterilization or radiation steriliza-
tion
No compressed air
No water jets, compressed air, brushes, or
similar to be used on the sensor insert
No ultrasonic bath
Clean and disinfect the flow sensor in accord-
ance with the corresponding instructions for
use.
For disinfecting the flow sensor use only clean
disinfectant solutions.
WARNING
Risk of fire
Residual vapors of highly flammable disin-
fectants (e.g., alcohols) and deposits that
were not removed during reprocessing may
ignite when the flow sensor is in use.
Ensure particle-free cleaning and disinfec-
tion.
After disinfection, allow the flow sensor to
air for at least 30 minutes.
Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
cles.
Replace flow sensors when damaged,
soiled, or not particlefree.