Instructions for use Zeus Infinity Empowered SW 2.n 65
Assembly and preparation
Connecting anesthetic gas receiving system (AGS)
In combination with Dräger anesthesia workstations
and their modules, the anesthetic gas receiving sys-
tem meets the requirements of the general safety
standards (ISO 8835-3 and ISO 80601-2-13).
The anesthetic gas receiving system does not work
as a stand-alone system, but is used as one of
3 components of an anesthetic gas scavenging
system (AGSS).
1 Connect the transfer hose to the waste gas con-
nector (A) on Zeus IE and to the port of the re-
ceiving system (B).
2 Plug the scavenging hose onto the port (C) of
the receiving system.
3 Connect the plug of the scavenginghose (D) to
the terminal unit of the disposal system.
z Follow the instructions for use of the anesthetic
gas receiving system (AGS).
180
CAUTION
Increased occupational exposure
Do not deliver any anesthetic agent if the AGS is
not connected!
A
C
D
B
WARNING
Risk of patient injury
A vacuum can be created in the breathing sys-
tem and in the patient's lung if the side open-
ings on the receiving system are covered.
Always make sure that the side openings on
the receiving system are not blocked.