Alarms
170
Instructions for use Zeus Infinity Empowered SW 2.n
Arrhythmia
z Touch the Arrhythmia tab (A).
The following parameters can be configured de-
pending on the option used:
B Dual lead processing: Off/On.
C Arrhythmia monitoring:
– Adv. (optional)
– Basic
– Off
Specify the settings for Rate and Count when an
arrhythmia event is reported.
Setting the Rate
z Touch button, set the required value and con-
firm.
For ASY, VF and ARTF the rate cannot be
changed.
Setting the Count
z Touch button, set the required value and con-
firm.
Set ranges, see "Set ranges of the patient monitor-
ing alarm limits" on page 179
Anesthesia
z Touch the Anesth. tab (A).
The alarm limits of the following parameters can be
configured:
B inN2O (Inspiratory N2O content of the breathed
air)
C inHal (Inspiratory anesthetic agent concentra-
tion of the breathed air)
D inEnf (Inspiratory anesthetic agent concentra-
tion of the breathed air)
E inIso (Inspiratory anesthetic agent concentra-
tion of the breathed air)
F inSev (Inspiratory anesthetic agent concentra-
tion of the breathed air)
G inDes (Inspiratory anesthetic agent concentra-
tion of the breathed air)
H etCO
2 (Endexpiratory CO2 content of the
breathed air)
I inCO
2 (Inspiratory CO2 content of the breathed
air)
J FiO
2 (Inspiratory O2 content of the breathed air)
K MVe (Expiratory minute volume)
L V
Ti (Inspiratory tidal volume)
M Paw (airway pressure)
N Brody Factor
368
Alarms
A
V.tach
>=
V.run
>=
AIVR
<=
SVT
>=
Tach
>=
Brady
<=
Pause
PVC
B
C
369
WARNING
Risk of patient injury
The user must adjust maximum inspiratory VA
concentration to suit the patient or else there
is a risk of concentration levels being too high
or too low.