Technical data
400
Instructions for use Zeus Infinity Empowered SW 2.n
Ambient conditions
Temperature range for the device
Operation +15 to +35 °C
Storage / transportation –15 to +40 °C
Relative humidity
Operation 25 % to 85 %, no condensation
Storage / transportation 25 % to 85 %, no condensation
Air pressure
Operation 700 to 1060 hPa
Storage / transportation 500 to 1060 hPa
CO
2 content of ambient air
Operation 300 to 800 ppm
When transporting filled DIVA metering modules,
the specifications of the anesthetic agent manu-
facturer must be observed.
Only store and transport DIVA metering modules if
they are empty! (page 76).
Alarm tone sequence IEC
Sound pressure level L(A) of the alarm tones at the
workstation, measured in accordance with
IEC 60601-1-8
Alarm volume (high priority) Settable from approximately 52 dB(A) to 70 dB(A)
Alarm volume (medium priority) Settable from approximately 51 dB(A) to 69 dB(A)
Alarm volume (low priority) Settable from approximately 50 dB(A) to 68 dB(A)
Auxiliary alarm 55 dB(A) and 75 dB(A)
Mains supply failure alarm 55 dB(A) and 75 dB(A)