Instructions for use Zeus Infinity Empowered SW 2.n 153
Operation
Editing a drug
In the Drug library dialog window:
1 Select a drug from the list using the rotary knob.
2 Touch the Edit button.
3 Edit settings in the basic data and deliveries
and confirm with the rotary knob.
If you make changes to drugs which are stored in a
profile, the modified data from the drug library is
transferred to the profile.
Deleting a drug
In the Drug library dialog window:
1 Select a drug from the list using the rotary knob.
2 Touch the Delete button and confirm with rotary
knob.
The drug is deleted.
Drug library import/export
In the Drug library dialog window:
1 Touch the Export or Import button.
2 Insert USB storage media and follow the in-
structions in the dialog window, see page 149.
3 Confirm with rotary knob.
Module DPS – transport concept
If a Module DPS syringe pump which is switched on
is plugged into the IVDock, the syringe pump set-
tings are applied by Zeus IE, even if a new patient
is loaded.
CAUTION
Risk of patient injury
After transferring a drug library, check the drug
configurations to eliminate the risk of incorrect
settings.
NOTE
The basic configurations of the pumps may be
partially changed when connecting to Zeus IE.
The changes to the basic configurations are pre-
served even after disconnection from Zeus IE.
They can be manually reset on the pump if re-
quired (see instructions for use for the Module
DPS).
NOTE
The data for the manual bolus are not automati-
cally adopted in the trend display for IV data if:
a manual bolus is started via the syringe pump
while Zeus is still in General Standby
or
the syringe pump has not yet been remotely
operated by the Zeus.
Only when the pump has been confirmed on
Zeus, the data from manually started boluses will
be adopted into the trend memory.