Cleaning, disinfection, and sterilization
360
Instructions for use Zeus Infinity Empowered SW 2.n
Dismantling components
After each patient
After a change of patient, only the HME filter or the
filter must be replaced at the device end.
HME filter or filter (A):
z Pull the HME filter off the nozzle of the Y-piece
and dispose of according to the relevant instruc-
tions for use.
Observe hospital regulations for infectious patients!
Daily
Sample gas line (B):
z Disconnect sample line from the CO
2 water trap
and the filter on the Y-piece.
HME filter or filter (A):
z Pull the HME filter off the nozzle of the Y-piece
and dispose of according to the relevant instruc-
tions for use.
Y-piece (C):
z Remove Y-piece
Breathing hoses (D):
z Remove the breathing hoses from the nozzles
of the Y-piece and the breathing system.
z Do not damage the breathing hoses!
When attaching or removing the breathing
hoses always hold them at the connection
sleeve and not at the spiral ribbing! Otherwise
the spiral ribbing can become detached from
the sleeve.
Breathing hoses with damaged spiral ribbing
can be kinked and interrupt ventilation!
Breathing bag (E):
z Remove breathing bag and breathing bag hose.
046
A
B
C
D
D
E
B
016