EasyManua.ls Logo

Dräger Zeus Infinity Empowered - Application; Device Monitoring; Intended Use; Ventilation Modes

Dräger Zeus Infinity Empowered
453 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Application
20
Instructions for use Zeus Infinity Empowered SW 2.n
Intended use
Zeus IE enables all forms of general anesthesia (in-
halation anesthesia, balanced anesthesia and total
intravenous anesthesia) and regional anesthesia,
as well as the monitoring of regional anesthesia.
Anesthesia is conducted with automatic ventilation,
manual ventilation or spontaneous breathing with
an oxygen ratio of at least 25 %, using a mixture of
Air (medical compressed air) or nitrous oxide with
pure oxygen and additional delivery of volatile an-
esthetic agents and/or intravenous drugs.
Application
The Anesthesia workstation system Zeus IE is de-
signed for use in medical rooms, operating rooms
and induction and recovery rooms.
Applications range from adults to pediatric patients
and neonates.
Ventilation is accomplished on the patient through
a laryngeal mask, a mask, or an endotracheal tube.
The breathing system can be used either with par-
tial rebreathing (low flow or minimal flow) or with
complete rebreathing (closed system).
Non-rebreathing system with external fresh-gas
outlet, for manual ventilation for connection to Bain
or Magill system for example (optional).
Zeus IE is equipped with modules for drug meter-
ing, for patient and device monitoring and for venti-
lation.
Ventilation modes
Volume-controlled ventilation at constant inspira-
tion flow Volume Control
Volume-controlled ventilation with decelerating in-
spiratory flow Volume Control AutoFlow;
synchronization of Sync.(VC) and Pressure Sup-
port breaths if required.
Pressure-controlled ventilation Pressure Control;
synchronization of Sync.(PC) and Pressure Sup-
port breaths if required.
Pressure-supported ventilation at constant positive
airway pressure Pressure Support
Manual ventilation/spontaneous breathing with and
without CPAP MAN/SPON
Device monitoring
The following are displayed as measured val-
ues:
Airway pressure PIP
Airway pressure Plat
Airway pressure PEEP
Airway pressure Mean
Minute volume MV
Tidal volume VT
Respiratory rate RR
CO
2, inspiratory and expiratory
O
2, inspiratory and expiratory
N2O, inspiratory and expiratory
Anesthetic agent concentration, inspiratory and
expiratory

Table of Contents