Fault – Cause – Remedy
250
Instructions for use Zeus Infinity Empowered SW 2.n
Failure behavior
In the event of an alarm and device-related failures,
the corresponding alarm message appears in the
alarm message field (A).
Power outage
If the power goes out, Zeus IE automatically
switches over to the integrated uninterruptible
power supply (UPS).
With a fully charged battery, operation is main-
tained for at least 30 minutes.
The Mains supply failure message appears in the
alarm message field.
The remaining battery capacity is displayed in per-
cent in the system field (B).
Example: 82 %
If the battery capacity is low, one of the following
messages appears:
– Battery low: Battery capacity <20 %
Or
– Battery very low: Battery capacity <10 %. The
device may switch off at any time!
In the event of a power failure and empty batteries,
Zeus IE enables manual ventilation and spontane-
ous breathing with 100% O
2.
– The ventilator, gas and anesthetic agent de-
livery as well as monitoring are inoperable.
– The DIVA metering modules are locked.
1 Unlock the O
2 rotary safety knob for emergency
O2 delivery by pressing it and set it to the re-
quired O2 flow. Range 0 to 12 L/min. Use pres-
sure gauge to monitor the pressure.
During this phase, anesthetic agent delivery is not
possible!
2 Ventilate the patient manually!
3 Recharge the empty battery by plugging in the
power cable for at least 10 hours.
301
1:35
A
B
CAUTION
Risk of patient injury
The Basis (IVDock) is no longer supplied with
power. Each Module DPS is powered by its own
battery. Pay attention to the battery warning! Infu-
sion is automatically stopped when the battery is
empty. Charge the battery as soon as possible.