Fault – Cause – Remedy
254
Instructions for use Zeus Infinity Empowered SW 2.n
After the Reboot:
Safety settings are activated, i.e. all the settings
are retained.
Exception for gas delivery: delivers 0.5 L/min of
O
2 without anesthetic agent.
To acknowledge the Safety settings active alarm:
z Press the Audio paused key.
Screen error
The screen does not respond to operation. It has
failed or the display is faulty.
The screen can be rebooted if necessary:
1 Touch and hold the On/Off button for a least
4 seconds.
The screen turns off.
2 After the screen turns off, touch the On/Off but-
ton again.
Reboot: Zeus IE performs a warm start of the
screen which takes approx. 3 minutes. Monitoring
is unavailable during this time.
Ventilation continues (fresh-gas mode at
0.5 L/min; with optional ext. FG outlet: 3 L/min).
Check ventilation with manual pressure gauge.
Perform manual ventilation if necessary.
After the Reboot:
Safety settings are activated, i.e. all the settings
are retained.
Exception for gas delivery: delivers 0.5 L/min of
O
2 without anesthetic agent.
To acknowledge the Safety settings active alarm:
z Press the Audio paused key.
In the event of device failure
If the device no longer responds or if the screen
fails, the following must be observed:
Anesth. Ventilator Off means:
Ventilation: Manual/Spontaneous
Gas delivery: 0.5 L/min of O
2
Anesthetic agent delivery: Off
1 Press the rotary safety knob O2 (B) to move it
out.
2 Turn the rotary safety knob O
2 (B) and adjust
the O2 flow.
3 If necessary, fill the breathing bag via O
2 flush
O
2+.
Contact DrägerService and use a substitute an-
esthesia machine.
CAUTION
Risk of patient injury
Anesthetic gas is not delivered.
CAUTION
Based on the mechanical airway manometer,
check whether ventilation continues. Manually
ventilate the patient if necessary!
If necessary, press the rotary safety knob Anesth.
Ventilator Off to move it out and activate by turn-
ing clockwise to switch the ventilation and gas de-
livery to manual emergency operation.
Ensure anesthesia. No anesthetic gases are de-
livered. Patient risk!