Instructions for use Zeus Infinity Empowered SW 2.n 363
Cleaning, disinfection, and sterilization
Drägersorb CLIC disposable absorber
1 Press the unlock button (B): the holder flips
open.
2 Slide the disposable absorber (D) out of the
holder.
3 Dispose of the used absorber, observe the
Drägersorb CLIC instructions for use, see also
page 67.
Breathing system
Remove the breathing system
1 Unlock both lateral twist locks (A).
2 Pull off the breathing system by the handle (B)
upwards.
To achieve optimum drying, sterilization at a
maximum temperature of 134 °C
(maximum 10 min) is possible after removal.
3 Unscrew nozzles.
4 Remove both SpiroLife flow sensors (C).
5 Remove the APL valve (D) by turning the
knurled nut (E) counterclockwise.
068
B
D
019177007
A
B
C
D
E