Instructions for use Zeus Infinity Empowered SW 2.n 423
Technical data
Cardiac output (C.O.)
Parameter display Cardiac output, blood temperature, injectate tem-
perature
Measuring principle Thermodilution
Measuring range Cardiac output: 0.5 to 20 L/min
Blood temperature: 25 °C to 43 °C
Injectate temperature: –5 °C to +30 °C
Accuracy Cardiac output: ±5 % (with 0 °C injectate)
Injectate temperature: ±0.25 °C
Electrical specifications Type CF as per IEC 60601-1
Defibrillation protection according to IEC 60601-1A2
Pulse oximetry (SpO2)
Parameter display Saturation (% SpO2), pulse rate
Absorption spectrophotometry
SpO
2: 70 to 100 %
Pulse rate: 30 to 250 bpm
70 to 100 %
0 to 100 %
2 seconds
30 seconds (in the event of artefact or other errors)
Measuring principle
Measuring range
Calibration range
Display range
Display update period
Maximum update delay