Assembly and preparation
88
Instructions for use Zeus Infinity Empowered SW 2.n
Emptying the DIVA metering module
1 Attach an empty anesthetic agent bottle.
2 Turn the DIVA metering module with the anes-
thetic agent bottle attached upside down and
push.
3 After emptying in this position, touch the unlock
button and remove the anesthetic agent bottle.
4 Turn the DIVA metering module upright and
close it with the locking cover.
5 Close the anesthetic agent bottle.
066