Instructions for use Zeus Infinity Empowered SW 2.n 255
Fault – Cause – Remedy
Failure behavior DPS
If Zeus IE is unable to operate a Module DPS sy-
ringe pump, the reasons can be as follows:
Module DPS is not a Dräger syringe pump and
has a software version which is not compatible.
Zeus IE cannot operate it. The syringe pump
continues to run.
Incorrect configuration, see page 140.
z Observe Module DPS instructions for use.
A syringe pump which has stopped and is then or
already connected to Zeus IE generates the alarm
message Infusion 1 stopped on Zeus IE after one
minute.
Messages Zeus IE
The alarm messages are displayed in the message
field of the header bar in hierarchical order.
Different background colors indicate the priority lev-
els of the alarms.
Zeus IE divides alarm messages by priority into
three categories:
To classify the alarms within an alarm category, in-
ternal priority figures are indicated in the following
table. The most critical alarm is given the number
255. Lower numbers indicate a lower alarm priority.
In the tables for Alarm info and Alarm log,
see page 161, the priority of the alarm messages is
also indicated by exclamation points.
Up to four high or medium-priority alarms from the
following four alarm areas can be displayed simul-
taneously:
Anesthesia
Hemodynamics (optional)
Ventilation
Infusion (optional)
From each alarm area at least one high-priority
alarm is displayed, given that there is an active
alarm.
Alarms are displayed according to priority, down-
wards from top to bottom and from left to right.
Low-priority alarms are only displayed in the right-
hand field of the alarm message field, the maximum
being four alarms.
The following table lists the alarm messages in al-
phabetical order. If an alarm occurs, the table helps
identify the cause and to quickly remedy by means
of the provided corrective actions. The possible
causes and remedial measures should be looked
through in the order they are listed until the alarm is
resolved.
The alarm messages with cause and remedy are
also displayed in the Alarms >Alarm info dialog
window.
Warning (red) High priority
Caution (yellow) Medium priority
Note (cyan) Low priority