Instructions for use Zeus Infinity Empowered SW 2.n 61
Assembly and preparation
Establishing potential equalization
Differences in electrical potential between devices
can be reduced by potential equalization.
Potential equalization does not replace the protec-
tive ground connection.
During operation, the potential equalization con-
nectors must be readily accessible and the connec-
tion must be able to be disconnected without the
use of tools.
1 Connect the potential equalization cable to the
potential equalization pin (A) on the device.
2 Connect the potential equalization cable to a
potential equalization connector of the hospital
(e.g., wall, ceiling supply unit, operating table).
Connecting the attachments
z Connect the attachments to the auxiliary sock-
ets on the back.
015
A
NOTE
If the mains power fails, devices connected to the
auxiliary power sockets are not supplied from the
uninterruptible power supply.
Do not connect any life-supporting devices to the
auxiliary power sockets of the anesthesia work-
station.
Ensure an alternative power supply for connected
devices.
CAUTION
Patient hazard due to risk of electric shock
Observe the maximum current input of the auxil-
iary devices (see relevant instructions for use).
WARNING
Connection of devices to the auxiliary sockets
of the anesthesia device can cause an in-
crease in the entire discharge current beyond
the permissible value if the protective conduc-
tor of a device fails. Therefore when connect-
ing devices to the auxiliary sockets, the total
leakage current has to be checked according
to IEC 60601-1-1.
When the permissible values for the entire dis-
charge current are exceeded, devices must
not be connected to the auxiliary sockets of
the anesthesia device, but to a separate wall
socket.
If this is ignored, it can put the patient and
user at risk.