Instructions for use Zeus Infinity Empowered SW 2.n 367
Cleaning, disinfection, and sterilization
Reprocessing methods
Classification of medical devices
For reprocessing, the medical devices and their
components are classified according to their type of
application and the resulting risks:
Non-critical medical devices:
Surfaces accessible to users and patients, e.g.,
device surfaces, cables
Semi-critical medical devices:
parts carrying breathing gas, e.g., breathing
hoses, masks
Testing of procedures and agents
The cleaning, disinfection, and sterilization of med-
ical devices were tested using the following proce-
dures and agents. The following agents showed
good material compatibility and effectiveness at the
time of the test:
Uncritical medical devices
Manual disinfection with simultaneous cleaning:
Incidin Extra N from Ecolab
Buraton 10F from Schülke & Mayr (for use on
the monitor only)
Semi-critical medical devices
Manual cleaning:
Neodisher FA, Neodisher Medizym from
Dr. Weigert
Manual disinfection:
Gigasept FF from Schülke & Mayr
Machine cleaning:
Neodisher FA, Neodisher Medizym from
Dr. Weigert
Machine disinfection:
Thermal, 93 °C (199.4 °F) for 10 minutes
Sterilization:
Hot steam, 134 °C (273.2 °F) for 5 minutes
WARNING
Risk of infection
Use validated reprocessing procedures when
reprocessing the device and accessories.