Startup
94
Instructions for use Zeus Infinity Empowered SW 2.n
Switching on Zeus IE
z Touch the On/Off button (A).
Manual ventilation and O
2 emergency delivery are
possible at all times.
Anesthetic agent is not delivered yet!
The initial screen appears after switching the de-
vice on.
The start may be delayed due to database mainte-
nance.
The device switches to the self test mode automat-
ically. The Test details > Checklist dialog window
appears.
050
WARNING
Perform the self test once a day in order to
check the operational reliability of the device
and hence minimize the risk of exposing the
patient to a hazard!
618
O2
Air N2OO2 Air N2O
A
310
Test details
Self test
Self test