Fault – Cause – Remedy
256
Instructions for use Zeus Infinity Empowered SW 2.n
Alarm – Cause – Remedy
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy
Note 150 2 mixed volatile
agents
A second volatile agent was
detected.
Switch to fresh gas mode
until the second gas is
washed out. Check the con-
tent of DIVA module.
Caution 060 3 mixed volatile
agents
A mixture of more than 2 vol-
atile agents detected!
Fresh-gas control: flush
with high fresh-gas flow.
Auto control: system
flushes automatically.
Check content of DIVA
module.
Caution 000 Acc. idioventr.
rhythm
Series of 3 or more PVCs
with a rate less than the VT
rate.
Check patient and take ap-
propriate action.
Caution 000 Action incomplete The last requested action did
not work.
Please try again.
Note 255 Agent low alarms
off
In Ventilator standby mode,
the lower alarm limits for
agent monitoring are disa-
bled.
Check measured values.
Note 050 Air central supply
high
Air central supply pressure
too high. Air pressure higher
than 8x100 kPa.
Check central Air supply.
System must not be used at
high pressures.
Note 200 Air central supply
low
Air central supply pressure is
low. Wall outlet is faulty. Air
plug is not connected to wall
outlet.
Check DIVA delivery.
Check O
2 concentration.
Remove Air plug from wall
outlet. Connect Air plug to
wall outlet.
Caution 200 Air cylinder empty Air cylinder empty or not
open (<20*100 kPa) and AIR
central supply is low.
Reduce fresh-gas flow, use
Auto control mode. Open or
exchange Air cylinder.
Check insp. O
2 concentra-
tion. Check central Air sup-
ply.
Note 225 Air cylinder low Air cylinder pressure is low
(20 – 55*100 kPa) and Air
supply is low.
Prepare exchange of Air
cylinder. Check Air supply.