Instructions for use Zeus Infinity Empowered SW 2.n 257
Fault – Cause – Remedy
Note 005 Air cylinder sensor
fail
Air high pressure sensor not
connected. Sensor failure.
Connect sensor. Check
sensor. Call DrägerService.
Caution 255 Air supply fail Air supply failed. Central
supply not connected. Com-
pressed Air hose kinked.
Check central Air supply.
Check connections. Open
optional Air cylinder. Pre-
pare IV Anesthesia.
Note 224 Air supply low Air supply low. Supply pres-
sure between 180 and
260 kPa.
Check central supply.
Caution 255 Air+N
2O delivery
cutoff
Air+N2O delivery locked.
Carrier gas was terminated
by protection system. Mixer
switches to 100 % O
2
Check insp. concentration.
Use IV anesthesia if neces-
sary. Call DrägerService.
Caution 000 Air+N
2O delivery
failure
Carrier gas delivery failure!
Mixer switches to 100 % O2
concentration!
Check inspiratory
concentration. Use IV
anesthesia if necessary.
Call DrägerService.
Caution 200 Air+N
2O delivery
failure
Carrier gas delivery failure.
Mixer switches to 100 % O2
concentration!
Check inspiratory concen-
trations. Use IV anesthesia
if necessary.
Call DrägerService.
Note 200 Alarms acknowl-
edged
Alarms have been acknowl-
edged by pressing the Audio
paused key. Alarm cause is
still valid.
Check acknowledged pa-
rameter.
Note 050 All alarms off All non-technical alarms are
suspended for 1min in order
to prevent false alarms.
This alarm will be deacti-
vated automatically.
Warning 255 Anesth. Ventilator
Off
Safety switch:Ventilator off
activated. Ventilator, mixer
and DIVA switch off.
Only use switch in an emer-
gency. Ventilate manually.
Open safety O
2 valve. Use
alternative anesthesia!
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy