Fault – Cause – Remedy
258
Instructions for use Zeus Infinity Empowered SW 2.n
Warning 180 Apnea No automatic ventilation. No
breath detected for 30 sec.
No spont/ass. ventilation: No
breath detected for 60 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Caution 250 Apnea No automatic ventilation: No
breath detected for 15 sec.
No spont/ass. ventilation: No
breath detected for 30 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Note 050 Apnea alarms off Apnea alarms are sus-
pended. To be used only dur-
ing intubation, suctioning
and manual ventilation.
Please pay special atten-
tion! To be used only during
intubation and suctioning
Warning 180 Apnea CO
2 No automatic ventilation: No
CO2 breath detected for
30 sec. No spont/ass. venti-
lation: No breath detected by
CO
2 for 60 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Caution 250 Apnea CO
2 No automatic ventilation: No
CO2 breath detected for
15 sec. No spont/ass. venti-
lation: No breath detected by
CO
2 for 30 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Note 175 Apnea flow No sufficient flow dynamics.
Flow sensor not calibrated.
To check and calibrate, per-
form self test or leak test.
Warning 180 Apnea flow No automatic ventilation: No
flow breath detected for
30 sec. No spont/ass. venti-
lation: No breath detected by
flow for 60 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Caution 250 Apnea flow No automatic ventilation: No
flow breath detected for
15 sec. No spont/ass. venti-
lation: No breath detected by
flow for 30 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy